AKPE-01M-03 EXP

AKPE-01M-03 EXP

Portable infrared breath alcohol analyzer with keyboard and printer built-in.

  • Automatic calibration of the measuring channel before each measurement.
  • Monitoring of the duration of the exhale.
  • Monitoring and recording of an interruption in the exhalation and an incomplete exhalation with a description of the irregularities on the display.
  • Measurement of the concentration of ethanol vapours in exhaled breath with a representation of the results on an alphanumeric 128x32 pixels LCD .
  • Involuntary sample selection for express analysis.
1 195 €

Product details AKPE-01M-03 EXP

AKPE-01M-03 portable brath alcohol analyser with independent and on-board power supply. Allows for express analysis with involuntary selection of an air sample through a funnel for a no-contact express analysis. The device operating principle is based on the spectrophotometric method of detecting ethanol vapours. The AKPE-01M-03 has absolute selectivity in relation to substances interfering with the detection of ethanol in exhaled breath. Storage of up to 4,000 measurement results with a date and time reference in the fiscal memory.

Functions:
• Storage of up to 4,000 measurement results with a date and time reference in the fiscal memory.
• Printing of the measurement results report on the built-in printer. The report contains the device's serial number, date and time of the measurements, date of the metrological verification, and the value of the measured concentration of ethanol vapours in exhaled air.
• Transfer of a block of measurement results to the computer.

Benefits:
• Objectivity of evaluation: prevents an inaccurate result in cases of recorded alcohol in the mouth cavity, reports an interruption in exhale or insufficient exhale force.
• Monitoring of ambient air: automatically monitors the alcohol contents in the ambient air and in the sample delivery system.
• Record printout: the record contains measurement results, date, time, device number, and record number. There is a protected memory of the results.
• Productivity: allows for up to 80 examinations per hour.
• Accuracy: does not require calibration throughout it service life, and maintains stable metrological characteristics without the use of calibration gas mixtures.

Specification AKPE-01M-03 EXP


Range of measurements of large ethanol vapour concentration in exhaled breath
From 0 to 1,500 mg/l

Range of readings
From 0 to 5,000 mg/l

Maximum permissible absolute error in the range of 0-0,200 mg/l
±0,020 mg/l

Maximum permissible relative error in a range higher than 0,200 - 1,500 mg/l
±10%

Time of measurement after sample selection
No more than 5 s

Set-up time after turning on at the ambient air temperature (20 ± 5) °С
No more than 3,5 min

Set-up time after measurement at the ambient air temperature (20 ± 5) °С
No more than 20 s

Time interval of analyser operation without adjustments to readings, no less than
12 months


Power supply
- On-board network (12.6±2) V - Built-in rechargeable battery - Network supply and charge (5±0,25) V

Number of measurements in analysers without recharging the rechargeable battery
No less than 250

Range of ambient air temperature
From -10 to +40 °С

Relative humidity of ambient air at a temperature of 25 °C
No more than 98%

Range of atmospheric pressure
From 66,6 to 106,6 kPa

Dimensional specifications, no more than
223х78х68 mm

Weight, no more than
0,5 kg

Power consumption, VA, no more:-in warm-up mode-in operating mode
102,5